MAINTENANCE OF TURBINES.

DOWNLOAD PDF

PREPARATION

Preparation

A.REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing

B.PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

Cleaning_A

A.CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B

B.REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C

C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

C’.CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION

Disinfectant-Tissue-spray_ATissue_rubbing_A

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

Thermal-disinfector

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

Lubrification_A1Lubrification_A2

A.LUBRICATION WITH LUBRIMED

OR

Lubrification_B1Support_B

B.LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 23 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

B’.INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION

Sterilization

A.PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer

B.STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

  • Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated

Posted on Leave a comment

Catalogues

Pdf IconContra-angles and handpieces 2424 kb DOWNLOAD
Pdf IconEVO.15 5811 kb DOWNLOAD
Pdf IconImplantology systems 4608 kb DOWNLOAD
Pdf IconLaboratory 121 kb DOWNLOAD
Pdf IconLubricare flyer 636 kb DOWNLOAD
Pdf IconMC2 + MC3 Micromotors 2759 kb DOWNLOAD
Pdf IconMCX MICROMOTOR 1373 kb DOWNLOAD
Pdf IconMX2 LED micromotor + MX2 System + DMX2 1999 kb DOWNLOAD
Pdf IconOptima 1602 kb DOWNLOAD
Pdf IconSystems and materials for dental laboratories 1337 kb DOWNLOAD
Pdf IconTornado 6352 kb DOWNLOAD
Pdf IconTurbines 2909 kb DOWNLOAD
Posted on Leave a comment

1600908-001

Latest edition 2100277 | Printed date: 02.18
English Pdf Icon2100277-0003_TORNADO_A6_ENG-20180215_FINAL3.pdf 4431 kb DOWNLOAD
French Pdf Icon2100277-0003_TORNADO_A6_FRA-20180215_FINAL1.pdf 4435 kb DOWNLOAD
German Pdf Icon2100277-0003_TORNADO_A6_DEU-20180215-FINAL.pdf 4436 kb DOWNLOAD
Italian Pdf Icon2100277-0003_TORNADO_A6_ITA-20180215-FINAL1.pdf 4433 kb DOWNLOAD
Japanese Pdf Icon2100277-0003_TORNADO_A6_JPN-20180223-FINAL3.pdf 4584 kb DOWNLOAD
Portugese Pdf Icon2100277-0003_TORNADO_A6_PRT-20180214_FINAL1.pdf 4436 kb DOWNLOAD
Russian Pdf Icon2100277-0003_TORNADO_A6_RUS-20180213-FINAL1.pdf 4719 kb DOWNLOAD
Spanish Pdf Icon2100277-0003_TORNADO_A6_ESP-20180216-FINAL1.pdf 4436 kb DOWNLOAD
Edition 2100277 | Printed date: 05.17
Czech Pdf Icon2100277-0000_TORNADO_A6_CZE-20170503_FINALVERSION.PDF 2620 kb DOWNLOAD
Edition 2100277 | Printed date: 02.15
English Pdf Icon2100277-0000_TORNADO_A6_ENG-20151006.pdf 2199 kb DOWNLOAD
French Pdf Icon2100277-0000_TORNADO_A6_FRA.pdf 2302 kb DOWNLOAD
German Pdf Icon2100277-0000_TORNADO_A6_DE.pdf 2327 kb DOWNLOAD
Italian Pdf Icon2100277-0000_TORNADO_A6_ITA.pdf 2218 kb DOWNLOAD
Japanese Pdf Icon2100277-0000_TORNADO_A6_JPN.pdf 2332 kb DOWNLOAD
Portugese Pdf Icon2100277-0000_TORNADO_A6_PRT.pdf 2935 kb DOWNLOAD
Russian Pdf Icon2100277-0000_TORNADO_A6_RUS-20151006.pdf 1655 kb DOWNLOAD
Spanish Pdf Icon2100277-0000_TORNADO_A6_ESP.pdf 2119 kb DOWNLOAD