MAINTENANCE OF TURBINES.

DOWNLOAD PDF

PREPARATION

Preparation

A.REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing

B.PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

Cleaning_A

A.CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B

B.REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C

C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

C’.CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION

Disinfectant-Tissue-spray_ATissue_rubbing_A

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

Thermal-disinfector

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

Lubrification_A1Lubrification_A2

A.LUBRICATION WITH LUBRIMED

OR

Lubrification_B1Support_B

B.LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 23 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

B’.INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION

Sterilization

A.PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer

B.STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

  • Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated